CamLoc Clamp

ProductID: 800-0271
$325.00

The CamLoc Clamp is designed for use in the sterile field.

  • Designed for a wide-range of purposes; affixing most operating room table accessories to most OR table designs and brands
  • Attaches to the operating table rail at any location over multiple drapes and accepts accessories with round posts up to 5/8″D (1.6 cm) and flat mounting posts to 1″ x 3/8″D (2.5 cm x .95 cm)
  • Patient weight capacity is 500 lbs (227 kgs)

Copyright Notice 
All rights reserved. Any duplication, adaption or translation without prior written consent is prohibited unless otherwise provided for in the relevant copyright laws.

CamLoc Clamp
800-0271
800-0271-DEN
800-0271-EU
800-0271-JPN
800-0271-SWISS
800-0271-UK


Initial Release:
7/31/2012


Manufacturer
SchureMed 
(081001460)
452 Randolph Street, Abington, MA 02351 USA 
Toll Free
 
(888) 724-8763 | Ph (781) 982-7000 | Fax (781) 982-7001 | orders@schuremed.com


Authorized Manufacturer
Emergo Europe Prinsessegracht 20, 2514 AP The Hague, The Netherlands

INSTRUCTIONS FOR USE

Step 1.
Place clamp onto rail (as shown).

 


Step 2. 
Lock clamp by turning the stainless ball left or right (as shown).

Step 3.
Inset the accessory (as shown).

Step 4. 
Lock the accessory by turning the star knob to the right 
(as shown).

Intended Use
The intended use for the CamLoc Clamp is to hold accessories to the surgical table.

General Specifications

Device Dimensions (maximum)
  • Length: 4 3/4” +/- 0.5”
  • Width: 5” +/- 0.5”
  • Depth: 2 1/4” +/-  0.5” 
  • Device Weight: 2.0 +/-  0.5 lbs.
  • Attaches to the rail of an operating room table at any point on the rail
  • Single person installation
Overview of Components
The CAMLOC CLAMP is an operating room table clamp that will attach to the side rails of any table. 

See side rail dimensions below:
US: 0.374” X 1.122” (9.5 mm x 28.5 mm) PN# 800-0271
Denyer: 0.236” x 1.496” (6 mm x 38 mm) PN# 800-0271-DEN   
Europe: 0.394” x 0.984” (10 mm x 25 mm) PN# 800-0271-EU
Eschmann (UK): 0.236” x 1.260” (6mm x 32mm) PN# 0271-UK
Japan: 0.354” x 1.260” (9 mm x 32mm) PN# 800-0271-JPN
Swiss: 0.394” x 1.1181” (10mm x 30mm) PN# 800-0271-SWISS

General Information
  • Product not made with (NRL) Natural Rubber Latex
  • Device supports 500 lb. proportional patient load (6’4” tall patient per 99% human body model)
  • Product warranty covers product from manufacturing defects for period of 2 years
  • If damaged in shipping, call Customer Service at (888) 724-8763 or (781) 982-7000 to obtain Return Material Authorization (RMA) number. For product warranty issues, contact Customer Service.
  • Product is maintenance-free, check product condition before next use
  • Life of device is 5 years under normal use
  • Store device between –4 °F to +86°F 
Disposal
  1. General - Prevent infection by cleaning and disinfecting product before disposal
  2. Packaging - Dispose packaging material via household waste according to national requirements
  3. SchureMed accepts back used or retired products - or dispose of product in accordance with national requirements


Product Use Warnings
WARNING!

The maximum load placed on the product may not exceed the appropriate proportion  of a patient weighing 500 lbs. Restrictions result from the permissible overall load, among other things, of accessory rails when using operating tables with a low maximum load capacity. Be absolutely sure to follow the instructions for use for your Operating Table system.


WARNING!

Hazard resulting from incorrect use. Be absolutely sure to follow the instructions for use for your  Operating Table system.

CLEANING INSTRUCTIONS
Clean CamLoc Clamp thoroughly before using (before sterilization) and after each use. Recommended cleaning products: (All solutions should be used in accordance with manufacturer’s instructions)
  • Enzymatic neutral pH detergent solution
  • (Enzymatic cleaners are compatible with metal devices/instruments and are preferred)
    Non-enzyme, hydrogen peroxide based solution
  • Coverage Plus NPD One Step Cleaner, Disinfectant & Deodorizer Concentrate
Hand Cleaning
  • Scrub each piece with a soft, plastic scrub brush and a recommended cleaning solution (see above)
  • Rinse all parts thoroughly to remove any chemical residue
  • Wipe each part completely dry with a clean, lint-free cloth
Washer Cleaning
  • Put through washer/decontaminator according to manufacturer’s instruction
    (Select  cycle that does not included lubrication)
  • Prepare for sterilization (device must be cleaned before sterilization)
  • See sterilization instructions below

*Based on information provided by the Healthcare Infection Control Practices Advisory Committee (HICPAC), the Center for Disease Control (CDC), the National Center for Biotechnology Information (NCBI) and the U.S. National Library of Medicine.

STEAM STERILIZATION 

GRAVITY
Wrapped Instruments
Exposure Time at 121° C (250° F): 30 minutes
Exposure Time at 132° C (270° F): 15 minutes
Exposure Time at 135° C (275° F): 10 minutes
Drying Time: 15-30 minutes

Unwrapped Nonporous Items
Exposure Time at 132° C (270° F): minutes
Exposure Time at 135° C (275° F): minutes
Drying Time: minute

PRE-VACUUM
Wrapped Instruments
Exposure Time at 132° C (270° F): 4 minutes
Exposure Time at 135° C (275° F): minutes
Drying Time: 20-30 minutes

Unwrapped Nonporous Items
Exposure Time at 132° C (270° F): 3 minutes
Exposure Time at 135° C (275° F)3 minutes
Drying Time: N/A